PRODUCTS >> Laboratory >> Clinical Chemistry >> Equipment
Electrolyte Analyzer
HUMALYTE


FLEXIBILITY
  • 2 systems with 3 parameters

  • Na, K, Cl or Li instrument

  • Serum, Plasma, Urine

  • Whole blood

SIMPLICITY

  • Direct ISE

  • Clear user guidance

  • Built-in printer

  • On-line quality control

 

Easy set-up of instrument parameters


  

Stand-by mode to save reagents


 



 Specifications

Parameters

Sodium, Potassium, Chloride or Lithium

LCD Screen

40 chars, x 4 lines Transreflective with LED Back Light

Sample Type

Serum, Plasma,

Thermal Printer

32 characters per lines

 

Whole Blood (Vacutainer, Syringe)

RS232 Serial Interface

ASTM protocol

Secondary

Urine (Manual pre-dilution required)

Bar Code Reader Port

RS232 type bar code scanner

Operating Range and Imprecision:

Sample: Serum, Plasma, Whole Blood

On-Board QC

7 different controls

Sample: Serum, Plasma, Whole Blood
  

    With-In-RUN, n=20 Day-To-Day, n=20  
Analyte [mmol/L] CV% SD [mmol/L] CV% SD [mmol/L] [mmol/L]
Sodium 60-200 1.0 1.0 2.0 1.0 0.1
Potassium 1.1-10.0 1.5 0.15 3.0 0.15 0.01
Chloride 50-200 1.5 2.0 2.0 2.0 0.1
Lithium 0.1-5.0 2.0 0.03 3.0 0.03 0.01
  

On-Board Storage

Last 250 patient samples

Sample: Urine
  

    With-In-RUN, n=20 Day-To-Day, n=20  
Analyte [mmol/L] CV% SD [mmol/L] CV% SD [mmol/L] [mmol/L]
Sodium 10-400 3.0 1.0 4.0 1.5 0.1
Potassium 10-300 3.0 0.5 4.0 0.5 0.1
Chloride 20-400 4.0 2.0 5.0 2.5 0.1
Lithium --- --- --- --- --- ---
  

Data Power Protection

Time, date, last 2-point calibration parameters, last patient ID, setup                                                    parameters and QC data are stored in nonvolatile memory chips

Sample Volume

120 mkl blood, serum, plasma or 120 mkl urine

Power Equipments

100-120V, 50-60Hz / 200-240 V

Carryover

Less than 1% sample to sample

Regulatory Approvals

CE Mark, FDA 510k

Analysis time

45 seconds/sample

Environment 

Humidity 85% for temperatures up to 30°C 50 % relative humidity at 40°C

Calibration

Automatic 2-point calibration

User selectable intervals of 1-4 hours, 1-point calibration with each  sample, Idle in STANDBY MODE (i.e. no calibration)

Operating Temperature

15"C ... 40"C

 

 

Recommended Altitude

up to 2000 m

 

 

Dimension

41õ27õ24.5 cm

(H×W×D), 12 kg