Measurement
Compartment |
Sources of radiant energy - two LEDs for motility and spectrophotometry channels
Detector system - two photo detectors - Motility and Optical Density |
Sample type |
Fresh
Washed
Frozen
Post-vasectomy |
Specificity |
Concentration: 85%
Motility: 80%
Normal morphology (WHO): 65%
High-sensitivity test: 95% motile cells determination |
Sensitivity |
Concentration: 90%
Motility: 85%
Normal morphology (WHO): 85% |
Visualization Compartment |
White LED illumination system
CCD, 330 TV lines
Objective: Standard, x20
Signal Output: PAL standard
Zoom system for magnification transition between x300 and x500
Focus regulator |
Displays |
Operational backlight LCD (16 lines x 40 characters)
Video backlight LCD (8 x 10 cm ) |
Printer |
Built-in, Dot Matrix
58 mm paper, with 20 characters per line
Ribbon cassette |
Keypad |
Operational keys
Numeric buttons
Cursor buttons
Video control keys |
Interface |
Video output (RCA)
RS-232 cable plug-in |
Accessories for testing sample |
Measuring capillary: Disposable, plastic, loaded syringe for an easy aspiration of sample
Standard laboratory microscope slide: depth 20 microns, coverslip 22x22 mm |
Memory |
500 test results
500 QC-measurement |
Operation system |
Analysis time: Normal test - 75 seconds, High-sensitivity/Postvasectomy test - 9 minutes
(5 minutes in version 2.45 of the SQA-V software)
Software: controls all functions, provides measurement algorithm (according to WHO'99),
and provides visual and automatic test performance
Temperature for sample testing: Calibrated for use at room temperature. Motility tests will
degrade over time and will vary if the sample is heated
Input signal from motility channel: analog, up to 5V
Input signal from spectrophotometer channel: modulated (1 kHz), analog, up to 5V |
Quality control |
Internal: electronic self- test and auto-calibration
External: “QwikCheck™beads” quality control material for zero level and concentration |