Features :
- Open system
- Easy assay programming. Sample and reagent positions can be moved
around using the Windows® "drag and drop" feature
- All HUMAN and IMTEC application available and preprogrammed
- Use of original test bottles
- Time-management-module
- QC-module integrated
- Report generator, printout format according to user demands
- Self-monitoring mechanics and optics
- IAD filters ,4 filters installed, 405, 450, 545, 630
- Two syringe pumps, sized: 2.5mL and 50 µl, optimized for volume
ranges from 5 µl – 1,95 ml.
- Password protection
|
|
|
|
|
Dimensions |
86cm W x 51cm L x 40cm H, approx.
weight = 35kg |
Reagent and Sample Dispensing: |
Capabilities: |
predilutions, dispensing single or multiple reagents |
Pumps: |
two syringe pumps, sized: 2.5mL and 50 µl |
Probe: |
316 stainless steel, level sensing |
Minimum and Maximum Volume |
5µl – 1.95ml |
Maximum number of specimens: |
96 (including calibrators and controls) |
Maximum number of reagents: |
Depending on the rack combination
Reagents can also be programmed to go to the Sample rack |
Reaction vessel: |
Standard microwells, strips |
Instrument bottles: |
1L Priming bottle, system Liquid
2L Wash bottle w/ low volume warning sensor;
1L Rinse bottle (or 2nd wash) w/ low volume warning sensor
Waste Bottel |
Incubation, timing and temperature control: |
Thermal control: |
Plate/Well 25°C , 37°C , or ambient
Temperature controlled to 25°C providing the
ambient room temperature is below 25°C
Sample Rack is not temperature controlled |
|
|
Washing |
Wash Head |
8-probe, automatic prime and rinse |
Programs: |
Create and run user programmable protocols
(aspirate, dispense, soak). Can wash wells for re-use as applicable |
Reading |
Optical design: |
Reads absorbance in four simultaneous channels; NIST traceable calibration; user selects monochromatic or bi-chromatic results |
Light Source: |
Tungsten-Xenon lamp |
4 position filter wheel: |
405, 450, 545, 630 |
Interference filters: |
Long life, hard coat, ion-assisted deposition,
+/- 2nm, 10nm typical half bandpass |
Linear range: |
-0.2 to 3.0A |
Photometric Accuracy: |
± (1% of the reading +0.005A from 0 to 1.5A)
± (2% of the reading +0.005A from 1.5 to 3.0A) |
SOFTWARE |
Format: |
CD-Rom and Internet upgrades |
Operating Systems |
Windows® 98, 2000, ME, NT 4.0, or XP |
Minimum System: |
Pentium®/ 133 MHz, 64 MB RAM, SVGA monitor, Serial port or USB port with Serial
adapter |
Recommended system: |
PentiumII®/333 MHz, 128 MB RAM, CD
drive, Windows® 98 or above |
Secondary menu options: |
Create/edit protocols, import/export data,
etc., Control, Run, Setup |
Calculation modes: |
Absorbance, single standard, factor, fixed time kinetics, kinetics by standard or factor, multi-calibrator point-to-point, linear regressions, log-logit, cubic spline, and nonlinear regressions (curve fit). |
Self monitoring modes: |
Lamp, bottle volume, filters, pressure, vacuum, mechanical function, and more |
QC options |
Store control data, print Levey-Jennings or QC range plots, calculate SDs |
Serial port: |
RS232 output only, 19200 Baud,1 start bit, 8 data, 1 stop, no parity, no handshake, Serial cable provided |
POWER |
Voltage Range : |
100-250VAC |
Frequency Range : |
50-60Hz |
Power Maximum: |
160W |
Installation Category: |
CAT II
|
Environmental Conditions for Safe Operation: |
Mains supply voltage: |
Fluctuations not to exceed ±10% of the nominal voltage |
Altitude: |
Up to 2000 m |
Humidity: |
80% for temperatures up to 31°C decreasing
linearly to 50% humidity at 40°C |
Temperature: |
5°C to 40°C |
Operating Temperature: |
18- 35°C recommended |
Operating Humidity: |
Less than 85% recommended |
|
|
|
|
|
|
|