| External Unit |
|
| Display |
graphic display |
graphic display |
graphic display |
touch screen |
touch screen |
touch screen LCD |
| Dimensions of display (mm) |
70x36 |
70x36 |
70x36 |
120x89 |
120x89 |
171x128 |
| Screen resolution (dots) |
128x64 |
128x64 |
128x64 |
320x240 |
320x240 |
640x480 |
| Keyboard |
|
combined alphanumetric and functional (all ECGs) |
touch screen |
| Electrode contact control |
indication of contact of each electrode (all ECGs) |
| Interface |
RS232, USB reduction (S and M line), RS232 USB (L line) |
| Dimensions (mm) |
276x168x74 |
276x168x74 |
330x270x70 |
330x270x70 |
407x312x125 |
407x312x125 |
| Weight Approx. (kg) |
2 |
2 |
3.2 |
3.2 |
5.7 |
5.9 |
| Printer |
|
| Paper width (mm) |
58 |
58 |
112 |
112 |
210/A4 |
210/A4 |
| Paper type |
roll |
roll |
roll |
roll |
roll, fax, z-folded, A4-office paper |
| Print type |
therminal |
therminal |
therminal |
therminal |
therminal, exterminal laser or ink printer |
| Paper speed (mm/s) |
5, 10, 25, 50 (all ECGs) |
| Signal processing |
|
| Number of channels |
1/3 |
3,6 |
3/6,12 |
6,12 |
12 |
12 |
| Number of leads |
12 (all ECGs) |
| Displayed leads |
1 |
1 |
1 |
3 |
12 |
12 |
| Printed leads |
1/1,3 |
1, 3, 6 |
3/3, 6, 12 |
3, 6, 12 |
3, 6, 12 |
3, 6, 12 |
| Transmission in automatic mode |
real-time
synchronic |
real-time
synchronic |
real-time
synchronic |
real-time
synchronic |
synchronic |
synchronic |
| Pacemarker detection |
yes (all ECGs) |
| Sensivity (mm/mV) |
2,5, 5, 10, 20 (all ECGs) |
| Stored records (10 sec long strips) |
6 |
6 |
-/15 |
70 |
120 |
120 |
| Long ECG recorder
1 lead/2 leads (min) |
9/no |
9/no |
9/no |
24/12 |
10/10 |
10/10 |
| Digital Filters |
|
| Adaptable mains filter (Hz) |
50-60 (all ECGs) |
| Muscle tremor filters (Hz) |
35, 25 (all ECGs) |
| Baseline filters (Hz) |
|
0,05 (3.2s), 0.11 (1.5s), 0.25 (0.6s), 0.50 (0.3s), 1.50 (0.1s) (all ECGs) |
| |
splines |
splines |
splines |
|
|
|
| Automatic |
- |
- |
- |
auto-adaptive |
auto-adaptive |
auto-adaptive |
| User-defined profiles |
|
| No. of profiles for auto. and man. mode |
3 |
3 |
1/5 |
10 |
unlimited |
unlimited |
| No. of profiles for long ECG recorder |
3 |
3 |
-/5 |
10 |
unlimited |
unlimited |
| Extension of channels |
1-3-6/3-6 |
- |
3-12/- |
- |
- |
- |
| Averaging of ECG curves |
-/optional |
optional |
optional |
optional |
optional |
optional |
| Analysis |
-/optional |
optional |
optional |
optional |
optional |
optional |
| Interpritation in words |
-/optional |
optional |
optional |
optional |
optional |
optional |
| PC connection with software
BTL-08 Win |
-/optional |
optional |
optional |
optional |
optional |
optional |
| Ergometry with software
BTL-08 Win Ergo |
-/optional |
optional |
optional |
optional |
optional |
optional |
|
Technical specifications |
BTL-08 S line |
BTL- 08 M line |
BTL- 08 L line
|
| Mains supply |
115 V/230V, 50-60 Hz |
115 V/230V, 50-60 Hz |
100 V-230 V, 50-60 Hz |
| Frequency response |
0.04 Hz – 150 Hz |
| Digital resolution |
3.9 µV |
| A/D conversion |
13 bits |
| Sampling frequency |
2000 Hz/12 leads, 18000 Hz/1 lead |
| Dynamic range |
15.9 mV |
| Polarisation voltage |
± 400 mV |
| Max. consistent voltage |
± 5 V |
| Input impedence |
>20 MOhm |
| Common mode rejection |
> 100 dB |
| Safety standards |
IEC 601-1, IEC 601-2-25, IEC 601-1-2, IEC 601-1-4, ISO 14971 |
| Accumulator capacity |
1-2 hour continuous recording
10 m continuous printig |
3-4 hour continuous recording
15 m continuous printig |
2-3 hour continuous recording
35 m continuous printig |
Charging time
(if completely discharged) |
max. 3 hours |
max. 10 hours |
max. 6 hours |
| Protection class |
II according to ECG 536 |
|
| |
The BTL-08 ECG Holter is designed and manufactured using advanced technology which provides precise and trouble-free operation even in the most difficult operating conditions. The communication between the module and a PC is wireless for an easy and fast operation. The patient's comfort is ensured by the minimum dimensions and low weight of the unit.
The equipment stores continuous 24-hour ECG record from 2 or 3 leads into the internal FLASH memory. It records pulse rate values and ST-segment course and evaluates evaluates variability of the R-R interval. The system identifies eleven basic types of arrhythmia. The patient button is well accessible. The unit keeps the records in the memory even if the power supply is disconnected and enables multiple reading: thus any possibility of losing an examination is eliminated. The 3-channel unit also features pacemarker detection and QT analysis. |
|
2-channel Holter
24-hour continuous monitoring from 2 leads
- Pulse rate and ST segment analysis
- Evaluation of the R-R interval variability
- QRS complex classification and rhythm analysis
|
|
3-channel Holter
24-hour continuous monitoring from 3 leads
- Pulse rate, ST, and QT segment analysis
- Evaluation of the R-R interval variability
- QRS complex classification and rhythm analysis
- Pacemaker detection and analysis
- Quick connection to PC via USB / IRDA
|
|
|
|
| C602.000v300 |
BTL 08 ECG Holter |
2 channel ECG Holter.
Standard unit: basic unit, case for holding apparatus and accessories, 4 rechargeable batteries, charger, IRDA port, serial connection cable for IRDA port, 50 self-adhesive adult electrodes, portable case with 2 straps |
| C602.020v400 |
BTL 08 ECG Holter |
3 channel ECG Holter.
Standard unit: basic unit, case for holding apparatus and acc., 4 rechargeable batteries, charger, IrDA port on USB adapter, USB connection cable for IrDA port, CD with USB-IrDA driver,50 self-adhesive adult electrodes, portable case with 2 straps |
|
|
The BTL-08 ABPM is a non-invasive ambulatory blood pressure monitor. Its design and construction fully comply with the latest standards and requirements. The equipment measures blood pressure by the oscillometric method. The used algorithm provides high-quality examination even in difficult conditions. The patient's comfort is ensured by the state-of-the-art module with functional design. |
- 24, 36 or 48 hour continuous monitoring of blood pressure
- LCD display on patient unit
- Easily accessible patient button
- Wireless communication between module and PC
- Analysis of blood pressure and pulse rate
- Graphic and table interpretation
- A4 print
| C601.000v200 |
BTL 08 ABPM |
Blood pressure monitor system.
Standard unit: Basic unit, adult cuff, IRDA port with serial cable, portable case with strap |
| Time of blood pressure monitoring |
24, 36 and 48 hours |
| Method of blood pressure monitoring |
Oscillometric |
| Memory capacity |
600 position |
| Accuracy |
± 3 mmHg |
| Applicable cuffs |
Children, adults |
| Data transfer to PC |
Via infrared port |
| PC minimum requirements |
Pentium 200 MHz or higher, 800x600 resolution |
| System requirement |
Wimdows95/98/2000/NT |
| Mains supply |
Accumulators AA, 4x1.2V, 1500 mAh
alkaline batteries AA, 4x1.5V |
| Dimensions |
128x77x45 mm |
| Weight |
310g including batteries |
| Standards |
IEC 601-1, IEC 601-1-1, IEC 601-1-2, IEC 601-2-30 |
| Safety |
Class II according to IEC 536 |
|
| |
|
The BTL-08 Win is an easy-to use software ideal to store, monitor, analyze and print ECG records.
It makes an excellent option for all BTL electrocardiographs.
- User-friendly software for 12-channel ECG recording
- System of patient cards organized in user-made patient databases
- Archiving of patient data with ECG records
- Easy search and sorting of records
- Long ECG recording
|
Full and/or shortened diagnosis
Interpretation in words
Comparison of any two ECG curves
Digital filtration
Online help
Cost-effective A4 printing on standard computer printer |
| CB008.092v303 |
BTL-08 Win |
Standard unit: BTL-08 S3, PC interface serial cable, patient cable with banana plugs, clamps, chest electrodes, SW for Windows. |
|
| |
|
The software BTL-08 Win Ergo is an advanced 12-channel system for reliable stress test monitoring and analysis. The system can be used with any BTL-08 ECG. All data can be processed through a computer network. The system communicates with a range of ergometers, such as treadmill and selected bicycle ergometers. Create your own examination protocols, or use a number of pre-programmed protocols, such as Balke Standard, Chung, Low Level Chung, Bruce protocol and Naughton Protocol Ellestad. |
- User-friendly software for 12-channel ECG recording
- Communication with a range of treadmills and ergometers
- Pre-programmed protocols, such as Balke, Chung, Bruce and Naughton
- User designed protocols
- Full and/or shortened diagnosis
- Interpretation in words
- Online ECG averaging
- Digital filtration
- Online help
- A4 print
|
| CB008.099V303 |
BTL-08 Win Ergo |
Standard unit: ECG unit BTL-08 S3, patient cable, pack of self-adhesive disposable electrodes, interface cable, adapter, user manual, software for stress-test. |
|
| |
|
|
|
|
|
|