PRODUCTS >> Medicine >> Sterilization >> Autoclaves
Autoclave
Cliniklav® 25



Reliability

The great importance of deciding for a professional autoclave becomes obvious to all when an autoclave breaks down in a doctor’s practice. Work with the patients slows or comes to a halt, operations may have to be postponed, and patients may even have to go home without treatment.

With its selection of the best components, and with its experience of 50 years in specialisation in the field of sterilization, MELAG guarantees that your therapeutic work will smoothly flow. This is possible with the latest in microprocessor control, high-quality vacuum pumps, innovative steam generators, and top-quality stainless steel. This MELAG high-tech autoclave satisfies stipulations of European Standard EN 285 for large-scale sterilizers in hospitals.

Cost effectiveness

For their operation, vacuum autoclaves require demineralised water for the generation of sterilization steam. They also need tap water for cooling the vacuum pump and, of course, electric power.  In order to keep costs of these media as low as possible, modern cost effectiveness enjoyed high priority in the design of the MELAG Cliniklav® 25. It was possible to optimise the effectiveness of the vacuum pump to such a degree that it effectively functions with tap-water consumption lower than any levels possible until now. The MELAG steam generator is just as thrifty in its use of distilled water. MELAG has been awarded a much-admired patent for the design of this steam generator.

User friendlines

Sterilization-program diversity must not become an end in itself.  After all, selection of the wrong program, with aborted sterilization cycles and the required new starts, all cost time and money. The clearly organised display of the MELAG Cliniklav®25 shows the four main programs and the three options which do not depend on the program. Upon request, it is possible to store additional programs in the microprocessor of this autoclave.

Cost effectiveness

For their operation, vacuum autoclaves require demineralised water for the generation of sterilization steam. They also need tap water for cooling the vacuum pump and, of course, electric power.  In order to keep costs of these media as low as possible, modern cost effectiveness enjoyed high priority in the design of the MELAG Cliniklav® 25. It was possible to optimise the effectiveness of the vacuum pump to such a degree that it effectively functions with tap-water consumption lower than any levels possible until now. The MELAG steam generator is just as thrifty in its use of distilled water. MELAG has been awarded a much-admired patent for the design of this steam generator.

User friendlines

Sterilization-program diversity must not become an end in itself.  After all, selection of the wrong program, with aborted sterilization cycles and the required new starts, all cost time and money. The clearly organised display of the MELAG Cliniklav®25 shows the four main programs and the three options which do not depend on the program. Upon request, it is possible to store additional programs in the microprocessor of this autoclave.

Features:

  • 12-bit analogue/digital converter

  • Redundant process monitoring which is independent of the control system

  • PT 1000 temperature sensor for exact closed-loop control and monitoring

  • Cost-effective water-ring vacuum pump

  • Integrated steam generator with save-water function

  • RS 232 serial interface for connection to a PC in the doctor’s practice; or, with a dot-matrix printer which can be built in

  • Safety door contact switch

  • Flexible connection hoses make installation of the autoclave just as simple as a domestic washing machine

Installation requirements:

  • DN 15 cold-water line; minimum flow pressure = 2.5 bar

  • Drain of water is either by DN 40 wall connection, with MELAG Water Stop Valve (art. no. 1056); or, by a drain
    in the floor

  • Electrical ratings:  three-phase power (3 N AC 400 V / 50-60 Hz); fusegear = 3 x 16 A; maximum installed load = 10 kW

Dimensions

Depth x Width x Height

Weignt

Tabletop model

84 x 61,6 x 80 cm

150 kg

Interior space

65 x 32 x 32 cm

Lower cabinet

84 x 61,6 x 149,6 cm

50 kg